General Specifications:
|
Voltage Read and Source:
|
Source
|
0.000 to 20.000 VDC
|
Read
|
Isolated 0.000 to 30.000 VDC Non-isolated 0.000 to 20.000 VDC
|
Thermocouple mV:
|
Read and Source
|
-10.000 to +75.000 mV
|
Current (mA):
|
Source
|
0.000 to 24.000 mA
|
Read
|
Isolated 0.000 to 24.000 mA Non-isolated 0.000 to 24.000 mA
|
Frequency (1 to 20 V selectable amplitude):
|
CPM Source and Read
|
2.0 to 600.0 CPM
|
Hz Source and Read
|
1.0 to 1000.0 Hz
|
kHz Source and Read
|
1.0 to 10.00 kHz
|
Pulse (Source only; 1 to 20V selectable amplitude):
|
Pulses
|
1 to 30,000.0 2 CPM to 10 kHz
|
Ohms:
|
Source Read
|
5.0 to 4000 Ohms 0.00 to 4000.0 Ohms
|
Thermocouple Read and Source:
|
J Thermocouple K Thermocouple T Thermocouple E Thermocouple R Thermocouple S Thermocouple B Thermocouple C Thermocouple XK Thermocouple BP Thermocouple L Thermocouple U Thermocouple N Thermocouple
|
200.0 to +1200.0 °C -200.0 to +1370.0 °C -200.0 to +400.0 °C -200.0 to +950.0 °C -20.0 to +1750.0 °C -20.0 to +1750.0 °C +600.0 to +1800.0 °C 0 to +2316.0 °C -200.0 to +800.0 °C 0 to +2500.0 °C -200.0 to +900.0 °C -200.0 to +400.0 °C -200.0 to +1300.0 °C
|
RTD Read and Source:
|
Ni120 (672) Pt100 (385) Pt100 (3926) Pt100 (3916) Pt200 (385) Pt500 (385) Pt1000 (385) Cu10 YSI400 Cu50 Cu100 Pt385 Pt385-50
|
-80.0 to +260.0 °C -200.0 to +800.0 °C -200.0 to +630.0 °C -200.0 to +630.0 °C -200.0 to +630.0 °C -200.0 to +630.0 °C -200.0 to +630.0 °C -100.0 to +260.0 °C +15.00 to +50.00 °C -180 to +200 °C -180 to +200 °C -200 to +800.0 °C -200 to +800.0 °C
|
Environmental:
|
Operating Temperature
|
-10°C to +50°C
|
Storage Temperature
|
-20°C to +70°C
|
Stability
|
±0.005% of reading/°C outside of 23 °C ±5°C
|
Power Requirements
|
6 VDC; Batteries 4x AA; alkaline or optional rechargeable
|
Physical (includes rubber boot)
|
Dimensions 8.7"H x 4.2"W x 2.3"D (220.9 x 106.6 x 58.4 mm) Weight 30.5 ounces (863 gms)
|
Accessories Included
|
Test leads, 4-AA alkaline batteries, NIST Certificate, and instruction manual
|
Optional Accessories
|
Carrying Case, Model CC572
|
Accuracy:
|
Voltage2
|
±0.015% of reading, ±2 mV
|
Thermocouple
|
mV ±0.02% of reading, ±10 µV
|
Thermocouple Errors (in °C; add 0.2 for Cold Junction Compensation error
|
RTD Read and Source
|
Type J 0.2 °C Type K 0.3°C Type T 0.2 °C Type E 0.2 °C Type R 1.2 °C Type S 1.2 °C Type B 1.2 °C Type C 0.6°C Type XK 0.2 °C Type BP 0.9°C Type L 0.2 °C Type U 0.25 °C Type N 0.4°C
|
Ni120 (672) 0.2 °C Pt100 (385) 0.2 °C Pt100 (3926) 0.2 °C Pt100 (3916) 0.2 °C Pt200 (385) 0.8°C Pt500 (385) 0.4°C Pt1000 (385) 0.2 °C Cu10 1.4°C YSI400 0.1°C Cu50 0.4°C CU100 0.3°C Pt385-10 1.4°C Pt385-50 0.4°C
|
RTD IEX-range:
|
|
Read
|
Source
|
Current (mA)
|
±0.015% of reading, ±2 µA
|
±0.015% of reading, ±2 µA
|
CPM
|
±0.05% of reading
|
±1 LSD ±0.05% of reading
|
Hz
|
±0.05% of reading
|
±1 LSD ±0.05% of reading
|
kHz
|
±0.05% of reading
|
±1 LSD ±0.125% of reading
|
Ohms 400 Ohm Range
|
±0.025% of reading, ±0.05 Ω
|
±0.025% of reading, ±0.05 Ω
|
4000 Ohm Range
|
±0.025% of reading, ±0.5 Ω
|
±0.025% of reading, ±0.5 Ω
|